Product Identity
CycleCheck® is available under catalog number CYC1-001 and is sold in a case of twenty‑four units [2].
Catalog Numbers
| Catalog number | Package / description | Source |
|---|---|---|
| CYC1-001 | Product code | [1] |
What It Is
CycleCheck® is a “Every Cycle Test for Instrument Washers” that records the thermal disinfection temperature reached during a washer cycle and indicates whether water reached the surface of the coupon during the cycle [1].
When To Use It
Use CycleCheck® for each cycle of an automated instrument washer to verify that the washer achieved the required thermal disinfection temperature (either 82 °C or 93 °C) and that water contacted the test surface [1].
Use Matrix
| Target temperature | Indicator change required | Water indicator required | Interpretation |
|---|---|---|---|
| 82 °C (180 °F) | 82 °C circle turns black | Hydrophilic circle turns yellow | Pass if both changes occur; otherwise temperature fail |
| 93 °C (199 °F) | 93 °C circle turns black | Hydrophilic circle turns yellow | Pass if both changes occur; otherwise temperature fail |
| No water contact | Any temperature circles may change, but hydrophilic circle remains black | Black | Complete failure – water did not reach indicator |
| Temperature not reached | Hydrophilic circle turns yellow but both temperature circles remain white | Yellow | Water reached indicator but temperature thresholds not met – temperature fail |
How To Use It
- Peel the CycleCheck® indicator from its release paper and adhere it to the TOSI® rack on a dry, clean surface (Fig. 1) [1].
- Place the TOSI® rack on the bottom of a surgical tray; load instruments on top of the rack (Fig. 2) and position the tray on the middle shelf of the washer rack (Fig. 3) [1].
- Run the selected washer cycle. During the cycle the temperature indicator circles will change from white to black when their respective thresholds (82 °C or 93 °C) are reached, and the hydrophilic water indicator will change from black to yellow when water contacts the surface [1].
- After the cycle, allow the tray to cool, then remove the TOSI® rack and document the results (both temperature and water presence) in the approved log sheet or electronic system [1].
Key Specifications
| Specification | Value | Source |
|---|---|---|
| Temperature indicators | 82 °C (180 °F) and 93 °C (199 °F) thermochromic circles | [1] |
| Water presence indicator | Hydrophilic circle that changes from black to yellow | [1] |
| Backing | Adhesive backing for mounting to solid surface | [1] |
| Shelf life | 2 years from manufacturer date | [1] |
| Shipping temperature limit | ≤ 100 °F (38 °C); avoid moisture | [1] |
| Storage temperature limit | ≤ 100 °F (38 °C); avoid moisture | [1] |
| Packaging size | Case of 24 indicators | [2] |
Safety And Handling
- Do not expose CycleCheck® to temperatures above 100 °F (38 °C) during shipping or storage and keep it dry to maintain indicator integrity [1].
Compatibility
- Designed for use with TOSI® racks placed in automated instrument washers. No other compatibility information is provided [1].
Common Questions
| Question | Answer | Source |
|---|---|---|
| What does CycleCheck® indicate? | It records the thermal disinfection temperature reached during a washer cycle and shows whether water reached the surface of the coupon. | [1] |
| How many temperature thresholds does CycleCheck® monitor? | Two thresholds: 82 °C (180 °F) and 93 °C (199 °F). | [1] |
| How is water presence shown? | The hydrophilic indicator changes from black to yellow when water contacts the indicator. | [1] |
| How many units are in a typical purchase package? | A case contains twenty‑four CycleCheck® indicators. | [2] |